EMI/EMC Test and Measurement Print

: 3M Ferrite Tiled Shielded Semi - Anechoic Chamber :

Features :
    E-field Shielding better than 100 dB from 30 MHz.
Applications :
    Radiated emission and susceptibility of electrical or electronics product can be measured upto 6 GHz.
Users :
    Industries in and around Kolkata and Eastern India.
Location of facility :
    SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
: Radiated Emission Measurement Facility :

Service name :
  • Measurement of Radiated emission from electrical or electronic product as per CISPR 32,Ed.2 : 2015 FCC part 15 standard.
Features :
  • Measurement distance can be set 3M and Rx antenna height can be varied from 1M to 4M. 360º azimuth rotation of the EUT can be given.
Applications :
  • Electrical or electronics product with input current less than 20A.
Users :
  • Industries in and around Kolkata and Eastern India.
Location of facility :
  • SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
  • Facility is FCC listed. Radiated emission can be measured from 30MHz to 6 GHz.
: Radiated Susceptibility Test Facility :

Service name :
  • Radiated susceptibility test of electrical or electronic product as per IEC 61000-4-3,Ed.3.2:2010 standard.
Features :
  • Measurement distance can be set 3M and Field upto 10 V/M can be created on the EUT. Field uniformity better than 6dB for 75% or more test points.
Applications :
  • Measurement distance can be set 3M and Field upto 10 V/M can be created on the EUT. Field uniformity better than 6dB for 75% or more test points.
Users :
  • Railway vendors and other Industries in and around Kolkata and Eastern India.
Location of facility :
  • SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
  • Radiated susceptibility can be tested from 80MHz to 6 GHz and sine wave modulation at 1KHz with 80% modulation depth.
: Magnetic Field Immunity Test Facility :

Service name :
  • Magnetic field Immunity test of electrical or electronic product as per IEC 61000-4-8 standard.
Features :
  • Field Immunity can be tested at 50Hz. Severity level can be upto 300 A/m for continuous and upto 1000 A/m for short duration.
Applications :
  • Electronics product with CRT display with input current less than 20A.
Users :
  • Industries in and around Kolkata and Eastern India.
Location of facility :
  • SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
  • Immunity can be tested at power line frequency and field strength upto 5 Gauss.
: Conducted Emission Measurement Facility :

Service name :
  • Measurement of conducted emission from electrical or electronic product as per CISPR 11,22 FCC part 15 standard.
Features :
  • Conducted emission measurement of 3 phase EUT is possible and input current upto 25A.
Applications :
  • Electrical or electronics product.
Users :
  • Industries in and around Kolkata and Eastern India.
Location of facility :
  • SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
  • Conducted Emission can be measured from 150KHz to 30MHz and also as per FCC standard.
: Electro Static Discharge Test Facility :

Service name :
  • ESD test of electrical or electronic product as per IEC 61000-4-2 standard.
Features :
  • The severity level for air discharge is upto ±15 KV and contact discharge is upto ±8 KV. Pulse rise time is between 0.7nS to 1nS.
Applications :
  • Electronic equipment sensitive to static discharge.
Users :
  • Railway vendors and other Industries in and around Kolkata and Eastern India.
Location of facility :
  • SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
  • ESD test of EUT can be carried out upto ±15 KV.
: Electrical Fast Transient / Burst Test Facility :
Service name :
    EFT/Burst test of electrical or electronic product as per IEC 61000-4-4 standard.
Features :
    Pulse Rise time is 5 nS, Pulse width is 50 nS, Burst duration is 15 microsecond, Burst Period is 300 microsecond and Test voltage upto ± 4 KV
Applications :
    Electrical or electronic equipment with input current less than 16A.
Users :
    Railway vendors and other Industries in and around Kolkata and Eastern India.
Location of facility :
    SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
    EFT/Burst of EUT can be carried out upto ±4 KV and upto 5 KHz repetition rate.
: High Energy Surge Immunity Test Facility :
 Service name :
    High energy surge immunity test of electrical or electronic product as per IEC 61000-4-5 standard.
Features :
    Pulse Rise time is 1.2 microsecond, Pulse width is 50 microsecond and Test voltage upto ± 4 KV.
Applications :
    Electrical or electronic equipment with input current less than 16A.
Users :
    Railway vendors and other Industries in and around Kolkata and Eastern India.
Location of facility :
    SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
    Surge generator can deliver 1.2/50 microsecond voltage surge across high load resistance and 8/20 microsecond current surge into a short circuit.
: Harmonics Emission Measurement Facility :
 Service name :
    Measurement of Harmonic current emission from electrical or electronic product as per IEC 61000-3-2 standard.
Features :
    5KVA programmable source and automated measurement.
Applications :
    Electrical or electronic equipment with input current less than 16A.
Users :
    Industries in and around Kolkata and Eastern India.
Location of facility :
    SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
    Harmonic current emission from EUT can be carried out upto 16A input current.