Test & Measurement Facilities Print


Features :
  • E-field Shielding better than 100 dB from 30 MHz.
Applications :
  • Radiated emission and susceptibility of electrical or electronics product can be measured upto 1 GHz.
Users :
  • Industries in and around Kolkata and Eastern India.
Location of facility :
  • SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.

Service name :
  • Measurement of Radiated emission from electrical or electronic product as per CISPR 11,22 FCC part 15 standard.
Features :
  • Measurement distance can be set 3M and Rx antenna height can be varied from 1M to 4M. 360º azimuth rotation of the EUT can be given.
Applications :
  • Electrical or electronics product with input current less than 20A.
Users :
  • Industries in and around Kolkata and Eastern India.
Location of facility :
  • SAMEER Kolkata Centre, Sector V, Salt Lake Electronics Complex.
Other Detail :
  • Facility is FCC listed. Radiated emission can be measured from 30MHz to 1 GHz

Service name :
  • Radiated susceptibility test of electrical or electronic product as per IEC 61000-4-3 standard.
Features :
  • Measurement distance can be set 3M and Field upto 10 V/M can be created on the EUT. Field uniformity better than 6dB for 75% or more test points.